Welcome to Contral Instrument Services

Fluke 753 Documenting Process Calibrator

More Views

Fluke 753 Documenting Process Calibrator

Excl. VAT: £4,764.00 Incl. VAT: £5,716.80
Contral - Official Distributor
4764

The Fluke 753 has features we know you’ll love: increased accuracy, enhanced reliability, and is easier to read in dark environments.
Excl. VAT: £4,764.00 Incl. VAT: £5,716.80
Free shipping
Description

Details

Work smarter. Work faster.

 

The 753 is a power multifunction documenting calibrator that lets you download procedures, lists, and instructions created with software-or upload data for printing, archiving, and analysis.

 

  • Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments 
  • Source/simulate volts, mA, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters 
  • Power transmitters during test using loop supply with simultaneous mA measurement 
  • Measure/source pressure using any of 29 Fluke 700Pxx Pressure Modules 
  • Create and run automated as-found/as-left procedures to satisfy quality programs or regulations. Record and document results 
  • Holds up to a full week of downloaded procedures and calibration results. Use many features like autostep, custom units, user entered values during test, one-point and two-point switch testing, square root DP flow testing, programmable measurement delay etc. 
  • Easy to use 
  • Three year warranty 
  • Bright white dual display. Read both sourced and measured parameters simultaneously. 
  • Multi lingual interface 
  • Rechargeable Li-Ion battery for 10 hour uninterrupted use. Includes gas gauge. 
  • Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 1 ms. 
  • Delivered with DPC/Track Sample software. 
  • Compatible with many Asset Management software packages.
Specifications

Additional Information

Specification
Measurement Accuracy
Voltage DC
   1 Year  2 Years
100.000 mV        0.02%+0.005 mV  0.03%+0.005 mV
3.00000 V  0.02%+0.00005 V  0.03%+0.00005 V
30.0000 V  0.02%+0.0005 V  0.03%+0.0005 V
300.00 V  0.05%+0.05 V  0.07%+0.05 V
Voltage AC
Range 
40 to 500 Hz     
 Resolution  1 Year 2 Years 
3.000 V  0.001 V  0.5%+0.002 V  1.0%+0.004 V
30.00 V  0.01 V  0.5%+0.02 V  1.0%+0.04 V
300.0 V  0.1 V  0.5%+0.2 V  1.0%+0.2 V
Current DC
   1 Year  2 Years
30.000 mA          0.01% + 5 uA   0.015%+7 uA
110.00 mA  0.01% + 20 uA  0.015%+30 uA
Resistance
   1 Year  2 Years
10.000 Ω           
 0.05% + 50 mΩ  0.07%+70 mΩ
100.00 Ω  0.05% + 50 mΩ  0.07%+70 mΩ
1.0000 kΩ   0.05% + 500 mΩ  0.07%+0.5 Ω
10.000 kΩ  0.1% + 10 Ω  0.15%+15 Ω
Frequency
   Resolution  Accuracy (2 Years)
1.00 to 110.00 Hz  0.01 Hz  0.05 Hz
110.1 to 1100.0 Hz  0.1 Hz  0.5 Hz
1.101 to 11.000 kHz   0.001 kHz  0.005 kHz
11.01 to 50.00 kHz  0.01 kHz  0.05 kHz

 

Source Accuracy
Voltage DC
   1 Year  2 Years
100.000 mV          
 0.01%+0.005 mV  0.015%+0.005 mV
1.00000 V  0.01%+0.00005 V  0.015%+0.0005 V
15.0000 V  0.01%+0.0005 V  0.015%+0.0005 V
Current DC
   1 Year 2 Years 
22.000 mA (Source)  0.01%+ 0.003 mA  0.02%+ 0.003 mA
Current Sink (Simulate)  0.02% + 0.007 mA  0.04% + 0.007 mA
Resistance
   1 Year  2 Years
10.000 Ω              
 0.01% + 10 mΩ  0.015% + 15 mΩ
100.00 Ω  0.01% + 20 mΩ  0.015% + 30 mΩ
1.0000 kΩ  0.02% + 0.2 Ω  0.03% + 0.3 Ω
10.000 kΩ  0.02% + 3 Ω  0.03% + 5 Ω
Frequency
   2 Years
0.1 to 10.99 Hz  0.01 Hz
0.01 to 10.99 Hz  0.01 Hz
11.00 to 109.99 Hz  0.1 Hz
110.0 to 1099.9 Hz  0.1 Hz
1.100 to 21.999 kHz  0.002 kHz
22.000 to 50.000 kHz         
 0.005 kHz

Technical Data
Data Log Functions
Measure functions:  Voltage, current, resistance, frequency, temperature, pressure
Reading rate:  1, 2, 5, 10, 20, 30, or 60 readings/minute
Maximum record length:  8000 readings (7980 for 30 or 60 readings/minute)
Ramp Functions
Source functions:  Voltage, current, resistance, frequency, temperature
Rate:  4 steps/second
Trip detect:  Continuity or voltage (continuity detection not available when sourcing current)
Loop Power Function
Voltage:  Selectable, 26 V
Accuracy:  10%, 18 V minimum at 22 mA
Maximum current:  25 mA, short circuit protected
Maximum input voltage:  50 V DC
Step Functions
Source Functions  Voltage, current, resistance, frequency, temperature
Manual Step  Selectable step, change with arrow buttons
Autostep  Fully programmable for function, start delay, stepvalue, time per step, repeat

 

Environmental Specifications
Operating Temperature
   -10 °C to +50 °C
Storage Temperature
   -20 °C to +60 °C
Dust/water resistance
   Meets IP52, IEC 529
Operating Altitude
   3000 m above mean sea level (9842 ft)

 

Safety Specifications
Agency Approvals
   CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993

 

Mechanical & General Specifications
Size
   136 x 245 x 63 mm (5.4“ x 9.6 x 2.5)
Weight
   1.2 kg (2.7 lb)
Batteries
Internal Battery Pack Li-ion:  7.2V, 4400mAh, 30 Wh
Battery Life
   >8 hours typical
Battery Replacement
   Via snap-shut door without opening calibrator; no tools required
Side port connections
   Pressure module connector
   USB Connector to interface to your PC
   Connection for optional battery charger/eliminator
Data storage capacity
   1 week of calibration procedures results
90 day specifications
   The standard specification interval for the 750 Series are 1 and 2 years.
   Typical 90 day measurement and source accuracy can be estimated by
   dividing the one year "% of reading" or "%of output" specifications by 2.
   Floor specifications, expressed as "% of full scale" or
   "counts" or "ohms" remain constant.

 

Temperature, Resistance Temperature Detectors
Degrees or % of reading
Type (α) Range °C Measure °C1 Source current Source °C Allowable current2
  1 year 2 year   1 year 2 year  
100 Ω Pt (385) -200 to 100
100 to 800
0.07 °C
0.02 % + 0.05 °C
0.14 °C
0.04 % + 0.10 °C
1 mA 0.05 °C
0.0125 % + 0.04 °C
0.10 °C
0.025 % + 0.08 °C
0.1 mA to 10 mA
200 Ω Pt (385) -200 to 100
100 to 630
0.07 °C
0.02 % + 0.05 °C
0.14 °C
0.04 % + 0.10 °C
500 μA 0.06 °C
0.017 % + 0.05 °C
0.12 °C
0.034 % + 0.10 °C
0.1 mA to 1 mA
500 Ω Pt (385) -200 to 100
100 to 630
0.07 °C
0.02 % + 0.05 °C
0.14 °C
0.04 % + 0.10 °C
250 μA 0.06 °C
0.017% + 0.05 °C
0.12 °C
0.034 % + 0.10 °C
0.1 mA to 1 mA
1000 Ω Pt (385) -200 to 100
100 to 630
0.07 °C
0.02 % + 0.05 °C
0.14 °C
0.04% + 0.10 °C
150 μA 0.06 C
0.017 % + 0.05 °C
0.12 C
0.034 % + 0.10 °C
0.1 mA to 1 mA
100 Ω Pt (3916) -200 to 100
100 to 630
0.07 °C
0.02 % +0.05 °C
0.14 °C
0.04 % +0.10 °C
1 mA 0.05 °C
0.0125 % + 0.04 °C
0.10 °C
0.025 % + 0.08 °C
0.1 mA to 10 mA
100 Ω Pt (3926) -200 to 100
100 to 630
0.08 °C
0.02 % +0.06 °C
0.16 °C
0.04 % +0.12 °C
1 mA 0.05 °C
0.0125 % + 0.04 °C
0.10 °C
0.025 % + 0.08 °C
0.1 mA to 10 mA
10 Ω Cu (427) -100 to 260 0.2 °C 0.4 °C 3 mA 0.2 °C 0.4 °C 0.1 mA to 10 mA
120 Ω Ni (672) -80 to 260 0.1 °C 0.2 °C 1 mA 0.04 °C 0.08 °C 0.1 mA to 10 mA

1For two and three-wire RTD measurements, add 0.4°C to the specifications.
2Supports pulsed transmitters and PLCs with pulse times as short as 1 ms

Temperature, Thermocouples
Type Source °C Measure °C Source °C
    1 year 2 years 1 year 2 years
E -250 to -200 1.3 2.0 0.6 0.9
  -200 to -100 0.5 0.8 0.3 0.4
  -100 to 600 0.3 0.4 0.3 0.4
  600 to 1000 0.4 0.6 0.2 0.3
N -200 to -100 1.0 1.5 0.6 0.9
  -100 to 900 0.5 0.8 0.5 0.8
  900 to 1300 0.6 0.9 0.3 0.4
J -210 to -100 0.6 0.9 0.3 0.4
  -100 to 800 0.3 0.4 0.2 0.3
  800 to 1200 0.5 0.8 0.3 0.3
K -200 to -100 0.7 1.0 0.4 0.6
  -100 to 400 0.3 0.4 0.3 0.4
  400 to 1200 0.5 0.8 0.3 0.4
  1200 to 1372 0.7 1.0 0.3 0.4
T -250 to -200 1.7 2.5 0.9 1.4
  -200 to 0 0.6 0.9 0.4 0.6
  0 to 400 0.3 0.4 0.3 0.4
B 600 to 800 1.3 2.0 1.0 1.5
  800 to 1000 1.0 1.5 0.8 1.2
  1000 to 1820 0.9 1.3 0.8 1.2
R -20 to 0 2.3 2.8 1.2 1.8
  0 to 100 1.5 2.2 1.1 1.7
  100 to 1767 1.0 1.5 0.9 1.4
S -20 to 0 2.3 2.8 1.2 1.8
  0 to 200 1.5 2.1 1.1 1.7
  200 to 1400 0.9 1.4 0.9 1.4
  1400 to 1767 1.1 1.7 1.0 1.5
C 0 to 800 0.6 0.9 0.6 0.9
  800 to 1200 0.8 1.2 0.7 1.0
  1200 to 1800 1.1 1.6 0.9 1.4
  1800 to 2316 2.0 3.0 1.3 2.0
L -200 to -100 0.6 0.9 0.3 0.4
  -100 to 800 0.3 0.4 0.2 0.3
  800 to 900 0.5 0.8 0.2 0.3
U -200 to 0 0.6 0.9 0.4 0.6
0 to 600 0.3 0.4 0.3 0.4
BP 0 to 1000 1.0 1.5 0.4 0.6
  1000 to 2000 1.6 2.4 0.6 0.9
  2000 to 2500 2.0 3.0 0.8 1.2
XK -200 to 300 0.2 0.3 0.2 0.5
300 to 800 0.4 0.6 0.3 0.6
 

 

   
Downloads

There are no downloads